Accuracy:
5% of measured value + 0.2 mg/L (with standard solutions) NH4-N
Cable length:
10 m fixed cable at sensor, extendable to 100m
Calibration method:
Sensorcode (calibrated sensor cartridge) or 2-point in-line value and matrix correction
Controller compatibility:
SC200, SC1000, SC4500. All controllers sold separately
Data types:
Current output, relays and bus communication via SC controller
Detection limit:
0.2 mg/L NH4-N
Flow:
< 4 m/s max.
Immersion depth:
0.3 - 3 m
Includes:
Ammonium probe, sensor cartridge, user manual
Installation:
45 ° ± 15 ° (vertical in flow direction)
Length:
320 mm
Material:
Cartridge: Stainless steel (1.4571), PVC, POM, ABS, NBR
Measurement method:
Potentiometric ion-selective measurement
Measuring interval:
Continuous
Measuring range:
0 - 1000 mg/L NH4-N
Measuring uncertainty:
±5% of the measured value + 0.2 mg/L (with standard solution) NH4-N
0.2 mg/L NH4-N (with standard solutions)
Operating temperature range:
Air: -20 - 45 °C (-4 - 113 °F)
Parameter:
NH4-N
pH Range:
5 - 9 pH
Process connection:
1 inch thread
Protection class:
IP 68
Repeatability:
±5 % of the measured value (with standard solution) NH4-N
Response time:
< 3 min
Sample pressure:
0.3 bar max.
Sample temperature:
2 - 40 °C
Sensor cartridge:
With Cartrical plus technology: compact housing containing calibrated electrodes for ammonium, potassium, reference system and temperature sensor, all calibrated to each other; typical lifetime 12 months
Storage conditions:
5 - 40 °C Sensor cartridge;
-20 - 60 °C Sensor
Warranty:
24 months
Weight:
2.4 kg
Wetted Materials:
Probe: stainless steel 1.4571; ASA + PC, silicon; PVC and PU;
Sensor cartridge: PVC; POM; ABS; stainless steel 1.4571; NBR;
What's included?:
Ammonium probe, sensor cartridge, user manual