Accuracy:
± 5% or ± 0.04 mg/L (whichever is greater) from 0 - 5 mg/L Cl₂
± 10% from 5 - 10 mg/L Cl₂
Air purge:
Optional with 3/8-inch quick-connect fitting and tubing; 0.003 m3/min at 1.38 bar maximum
Alerts:
Low and high chlorine. Dirty cell. Low and high sample flow.
Certifications:
CE compliant with: EN 61326-1, CISPR 11, EN 50581
ACMA RCM
South Korea KC Certificate
North America: FCC Supplier’s Declaration of Conformance, IEC/EN 60529, ICES-003
Controller compatibility:
SC200, SC1000, SC4500. All controllers sold separately
Cycle Time:
2.5 minutes
Detection limit:
0.03 mg/L
Dimensions (H x W x D):
342 mm x 329 mm x 177 mm
Drain:
½-inch ID flexible hose
Enclosure waterproof rating:
IP66
Flow rate:
60 - 200 mL/min through the instrument
Inlet:
¼-inch OD polyethylene tube, quick-disconnect fitting
Inlet pressure:
0.3 - 5.2 bar supplied to Y-strainer
0.1 - 0.3 bar supplied to analyser
Installation Kit:
Pressure Regulator Installation Kit
Interferences:
Other oxidising agents such as bromide, chlorine dioxide, permanganate, and ozone will cause a positive interference. Hardness must not exceed 1,000 mg/L CaCO₃.
Light source:
LED, measurement at 510 nm, 1 cm light pathlength
Measuring range:
0 - 10 mg/L Cl₂
Mounting:
Wall mount
Operating humidity:
0 - 90% relative humidity, non-condensing
Operating temperature range:
5 - 40 °C
Parameter:
Chlorine, free; Chlorine, total residual
Power requirements (Voltage):
12 VDC, 400 mA maximum (supplied by the controller)
Reagent consumption:
0.5 L of buffer solution and 0.5 L of indicator solution in 31 days
Sample requirements:
Y-strainer filtration with 40-mesh screen or higher
Sample temperature:
5 - 40 °C
Warranty:
24 months
Weight:
4.1 kg
What's included?:
CL17sc Chlorine Analyser with Pressure Regulator Installation Kit, and European User Manual. User provides screws and/or anchors for mounting.
Hach SC Controller and Reagents are not included, they are sold separately.