Accuracy:
Free Chlorine: ±10 % of the reference test (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Free Chlorine: ±3 % of the reference test (DPD) at constant pH less than 7.2 (±0.2 pH unit)
Total Chlorine: ±10 % of the reference test (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Total Chlorine: ±20 % of the reference test (DPD) at stable pH greater than 8.5
Cable connection:
5 pin, M12 connector
Cable length:
1 m (between gateway and SC controller)
Calibration method:
1-point or 2-point (zero and slope) calibration
Certifications:
CE/ETL, EMC
Connection drain line:
1/2 in ID
Connections:
1/4 in OD
Controller:
Panel only
Controller compatibility:
SC controller platforms
Detection limit:
30 ppb (0.03 ppm)
Diameter:
25 mm
Dimensions:
mm x 25 mm (7.68 in x 0.98 in)
Sensor Only: 195
Dimensions (H x W x D):
483 mm x 495 mm x 151 mm
Fitting:
Metric
Flow rate:
30 - 50 L/h, optimal is 40 L/hour
Includes:
Free chlorine sensor, chlorine sensor block, SC gateway, stainless steel panel and pH flow cell for grab samples
Interferences:
Free Chlorine: Monochloramine, chlorine dioxide, ozone, and chalk deposits. Total Chlorine: Chlorine dioxide, ozone and chalk deposits
Length:
Sensor: 195 mm
Material:
Panel: Corrosion-resistant, fully-submersible (stainless steel, PVC, silicon rubber and polycarbonate)
Measurement method:
Reagentless, electrochemical, three-electrode amperometric system
Measuring range:
0 - 20 ppm Cl₂
Mounting:
Flat, vertical surface
Operating temperature range:
0 - 50 °C (assured temperature compensation up to 45 °C)
Parameter:
Free Chlorine
pH Range:
4 - 9
Power requirements (Voltage):
12 V DC ±10%, 100 mA maximum (supplied by SC controller)
Pressure range:
0.5 bar, no pressure impulses and/or vibrations
Repeatability:
30 or 3%, whichever is greater
Response time:
Free Chlorine: 140 s or less for 90% change (T90) at a stable temperature and pH
Total Chlorine: 100 s or less for 90% change (T90) at a stable temperature and pH
Sample temperature:
5 - 45 °C
Storage conditions:
-20 °C - 60 °C dry
Temperature compensation:
Internal temperature compensation in the sensor
Internal temperature compensation in the sensor
Warranty:
24 months
Weight:
Approximately 5.5 kg (12 lbs)
Panel and empty panel-mounted components only
What's included?:
Free chlorine sensor, chlorine sensor block, SC Gateway, stainless steel panel and pH flow cell for grab samples
Width:
495.3 mm