Ambient temperature:
10 - 30 °C (50 - 86°F) ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Analogue outputs:
Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard: 4, optional: 8
Automatic cleaning:
Yes; Fequency freely programmable: 6 hours, 12 hours, daily, weekly
Calibration:
Automatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced
Certifications:
CE, ETL certified to UL and CSA safety standards, UKCA
Cycle Time:
20 min (dilution + 5 min)
Demineralised water:
When needed for rinsing / dilution
Detection limit:
≤ 1 µg/L
Digital outputs:
Profibus DP or Modbus RTU
Dimensions (H x W x D):
688 mm x 460 mm x 340 mm
Drain:
Atmospheric pressure, vented, min. Ø 32 mm
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Flow rate:
100 - 300 mL/min
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
Interferences:
Ions like Nitrite > 5 mg/L, Sulphide > 100 mg/L and Sulphite. Thiocyanate will cause high results. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Material:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Measurement method:
Colorimetric measurement at 578 nm using chloramine T method, conform with standard method APHA 4500-CN €
Measuring range:
0.001 - 0.02 mg/L
0.002 - 0.05 mg/L
0.002 - 0.1 mg/L
0.005 - 0.2 mg/L
0.025 - 1 mg/L (with internal dilution)
0.050 - 2 mg/L (with internal dilution)
0.125 - 5 mg/L (with internal dilution)
0.250 - 10 mg/L (with internal dilution)
0.375 - 15 mg/L (with internal dilution)
0.500 - 20 mg/L(with internal dilution)
Number of sample streams:
Standard: 1
Optional accessory for 2, 4, or 8 sample streams
Output:
Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP
Parameter:
Cyanide, free
Power:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Precision:
Better than 2% full scale range for standard test solutions
Protection class:
IP44
Reagent Requirements:
Keep between 5 - 20 °C in use
Relays:
5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150)
Sample pressure:
By external overflow vessel (open to atmospheric pressure)
Sample quality:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample temperature:
10 - 30 °C
Validation:
Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
Warranty:
24 months
Weight:
Max. 35 kg
What's included?:
EZ1012sc Cyanide Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and 2 x empty 2.5L Reagent Containers with Fittings (for Buffer and Chloramine Solution) and 1 x empty 2.5L Glass Reagent Contaner with Fittings (for Colour Solution)