EZ1031sc Phosphate Colorimetric Analyser, Yellow Method, 1 stream, 4x mA Output

EZ1031sc Phosphate Colorimetric Analyser, Yellow Method, 1 stream, 4x mA Output
Product #: EZ1031.9800140T
Unit Price Contact Hach

Trusted Equipment for Critical Operations

The EZ sc Series combines robust engineering with high-quality materials to ensure reliability 24/7. Designed for long-term performance, EZ sc Series minimises breakdowns and supports sustained operational success.

Scalable Solutions for Growing Needs

Built to evolve alongside your needs, online analysers from the EZ sc Series offer adaptable parameters and easier integration. They ensure your facility remains future proof without requiring expensive upgrades. Easily adjust parameter testing ranges and expand sample streams to accommodate innovation and growth.

Real-Time Data for Faster Decisions

The EZ sc Series offers advanced connectivity and real-time monitoring, delivering immediate access to actionable insights. Facilities utilising this technology benefit from improved reaction times, empowering more informed decision-making and quicker responses during critical situations.

Simplified Troubleshooting with Expert Support

Built-in diagnostics combined with secure remote data access make the EZ sc Series simplify issue resolution. Supported by a highly responsive team of experts, it ensures uninterrupted functionality and minimises operational downtime. By addressing and identifying issues early, EZ sc Series of online analysers helps prevent costly disruptions and repairs while maintaining system efficiency.

Clear Workflows for Predictable Outcomes

Standardised tools and guided workflows ensure consistent implementation and operation. The EZ sc Series minimises miscommunication, sets clear expectations, and reduces process inconsistencies, driving greater efficiency.

What's included?

EZ1031sc Phosphate Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and 1 x empty 2.5L Reagent Container with Fittings (for Colour Solution)

Specifications

Ambient temperature: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Analogue outputs: Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard: 4, optional: 8
Automatic cleaning: Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
Calibration: Automatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced
Certifications: CE, ETL certified to UL and CSA safety standards, UKCA
Cycle Time: No Dilution
Default: 10 min.
Continuous: 9 min.

Internal Dilution
Default: 15 min.
Continuous: 12 min.
Demineralised water: Internal dilution option: Rinse water & Dilution water
Detection limit: No Dilution
0.02 - 1 mg/L PO₄-P: 0.02 mg/L
0.05 - 2.5 mg/L PO₄-P: 0.05 mg/L
0.05 - 5 mg/L PO₄-P: 0.05 mg/L
0.1 - 10 mg/L PO₄-P: 0.1 mg/L

Internal Dilution
0.5 - 50 mg/L PO₄-P: 0.5 mg/L
1 - 100 mg/L PO₄-P: 1 mg/L
2.5 - 250 mg/L PO₄-P: 2.5 mg/L
5 - 500 mg/L PO₄-P: 5 mg/L
7.5 - 750 mg/L PO₄-P: 7.5 mg/L
10 - 1000 mg/L PO₄-P: 10 mg/L
Digital outputs: Profibus DP or Modbus RTU
Dimensions (H x W x D): 688 mm x 460 mm x 340 mm
Drain: Atmospheric pressure, vented, min. Ø 32 mm
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Flow rate: 100 - 300 mL/min
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
Interferences: Positive interferences may be caused by arsenate (AsO₄³⁻) and, under heated conditions, by silica or silicate (SiO₄-)
Reducing substances such as sulfide (S²⁻), thiosulfate (S₂O₃²⁻), thiocyanate (SCN⁻), and excess molybdate (MoO₄²⁻) may interfere at high concentrations. The following ions do not cause interference at concentrations up to 1,000 mg/L: aluminium, ferric iron, magnesium, calcium, barium, strontium, lithium, sodium, potassium, ammonium, cadmium(II), manganese(II), lead(II), mercury(I), mercury(II), tin(II), copper, nickel, silver(I), uranium(IV), zirconium(IV), arsenite, bromide, carbonate, perchlorate, cyanide, iodate, nitrate, nitrite, sulfate, sulfite, pyrophosphate, tetraborate, selenate, benzoate, citrate, oxalate, lactate, tartrate, formate, and salicylate.
Large quantities of color and turbidity interfere. Fats, oil, proteins, surfactants and tar interfere.
Material: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Measurement method: Colorimetric measurement vanadates yellow method compliant with APHA 4500-P (C)
Measuring range: No Dilution
0.02 - 1 mg/L PO₄-P
0.05 - 2.5 mg/L PO₄-P
0.05 - 5 mg/L PO₄-P
0.1 - 10 mg/L PO₄-P

Internal Dilution
0.5 - 50 mg/L PO₄-P
1 - 100 mg/L PO₄-P
2.5 - 250 mg/L PO₄-P
5 - 500 mg/L PO₄-P
7.5 - 750 mg/L PO₄-P
10 - 1000 mg/L PO₄-P
Number of sample streams: Standard: 1
Optional accessory for 2, 4, or 8 sample streams
Output: Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP
Parameter: Phosphate (vanadate yellow)
Power: 100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Precision: Better than 2% full scale range for standard test solutions
Protection class: IP44
Reagent Requirements: Keep between 10 - 30 °C
Relays: 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150)
Sample pressure: By external overflow vessel (open to atmospheric pressure)
Sample quality: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample temperature: 5 - 30 °C
Validation: Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
Warranty: 24 months
Weight: Max. 35 kg
What's included?: EZ1031sc Phosphate Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and 1 x empty 2.5L Reagent Container with Fittings (for Colour Solution)