FP360sc PAH/Oil Fluorescence probe, 0-5000 ppb, Titanium Body, 10 m, with cleaning

FP360sc PAH/Oil Fluorescence probe, 0-5000 ppb, Titanium Body, 10 m, with cleaning
Product #: LXV441.99.22201
Unit Price Contact Hach

The right technology for the right price

Due to its unique combination of submersible probe design and UV fluorescence sensing technology, the FP 360 sc delivers the best technology to detect oil in water and is priced below competitive UV fluorescent instruments.

Minimal maintenance

The FP 360 sc has no tubes, pumps, or valves that can foul or require constant maintenance interventions. Maintenance is limited to occasional wiping of the sensor's measurement window, calibration once every two years, and Xenon lamp replacement every four years.

Reduced laboratory testing

While laboratory testing is the ultimate method of measuring oil in water, it is a long and complex process that requires special equipment and trained lab personnel. The FP 360 sc provides a cost-effective, continuous online monitoring solution to maintain process control and avoid oil contamination with minimal laboratory testing.

High sensitivity and selectivity

The FP 360 sc can detect and measure polycyclic aromatic hydrocarbons (PAHs) from 1.2 ppb to up to 5000 ppb (µg/L). This is approximately equivalent to a concentration of mineral oil between 0.1 to 150 ppm (mg/L). Furthermore, the FP 360 sc method of detection makes it impervious to interferences by turbid water or natural organic and biological matter that impact online light scattering, UV absorbance, and VIS fluorescence instruments.

One to eight sensors

The Hach Digital Controller Family can receive data from up to eight Hach digital sensors, including oil-in-water, suspended solids, turbidity, pH/ORP, dissolved oxygen, conductivity, ammonium, phosphate, SAC, and nitrate in any combination.

There’s no complicated wiring or set up procedures with the family of controllers. Just plug the sensor to any Hach digital controller and it’s ready to use because it’s “plug and play.”

What's included?

Oil-in-water probe, user manual

Specifications

Ambient temperature: -5 - 45 °C
Body Material: Titanium
Cable length: max. 40 m (130 ft) incl. extension cable
Calibration: Factory calibrated with UV fluorescence standard or process calibration with results of a grab sample analysis.
Detector: UV photodiode with interference filter; Compensation of daylight and flashlamp intensity fluctuations
Diameter: 68 mm
Excitation: Wavelength 254 nm
Housing material: Titanium
Length: 311 mm
Light source: Miniature xenon flashlamp with interference filter
Material: Housing: Titanium
Measurement method: UV fluorescence method for polycyclic aromatic hydrocarbons (PAH)
Measuring range: 0 - 5000 ppb (µg/L) (PAH) or 0.1 - 150 ppm (mg/L) (Oil)
pH value(s): ≥ 4
Pressure range: Max. 30 bar (measurement probe)
Reproducibility: 2.5 % of measured value at constant temperature
Response time: 10 s
Sample temperature: 1 - 40 °C
Sensor cleaning: No
Storage conditions: -40 °C - 60 °C
Warranty: 24 months
Weight: 1.8 kg
What's included?: Oil-in-water probe, user manual