Accuracy:
0 - 500 µg/L: ±1% or ±1 µg/L of reading, whichever is greater;
500 - 5000 µg/L: ±5%
Alarm:
Four relays output; type: not powered SPDT relays, each rated at
5 A resistive, 240 V AC maximum
Connection: 0.82 to 2.08 mm2 wire (18 to 14 AWG ), 0.82 mm2 stranded (18 AWG) recommended
Application:
Pure water / Power
Certifications:
CE (EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000)
KC (EN 61326-1: 2006)
C-tick (EN 61326-1: 2006)
cETLus (UL 61010-1: 2012; NEMA 250: 2003; CSA C22.2 No 61010-1: 2012)
Controller compatibility:
SC200, SC1000
Detection limit:
0.5 µg/L
Dimensions (H x W x D):
804 mm x 452 mm x 360 mm
Enclosure waterproof rating:
IP56 / NEMA 4X
Fitting:
Sample line and sample bypass drain: 6 mm (¼-in.)
Air purge inlet: 6 mm (¼-in.) Chemical and case drains: 9.5 mm (3/8-in.)
Flow rate:
55 - 300 mL/min
Grab sample:
Grab Sample In and Grab Sample Out capability
Light source:
Class 1M LED (light emitting diode) with a peak
wavelength of 810 nm
Manual languages:
English, French, Spanish, B. Portuguese, Chinese, Japanese,
Korean, Thai, German, Italian, C. Portuguese, Czech, Danish, Dutch, Polish, Swedish, Finnish, Bulgarian, Hungarian, Romanian, Lithuanian, Russian, Turkish, Slovak, Slovenian, Croatian, Greek, Estonian
Measuring principle:
Colorimetric
Measuring range:
0 - 5000 µg/L as SiO2
Mounting:
Wall, panel or table
Number of channels:
1, 2, 4, 6; programmable sequence
Operating humidity:
5 - 95 % relative humidity, non-condensing (indoor use only)
Operating temperature range:
5 - 45 °C
Output:
4 to 20 mA
Parameter:
Silica
Pollution degree:
2/II
Power requirements (Hz):
50/60 Hz
Power requirements (Voltage):
100 - 240 V AC, 24 V DC
Protection class:
I
Reagent consumption:
2 L of each reagent every 90 days with 15 minute cycle time
Relays:
Four; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC maximum
Repeatability:
± 0.5 µg/L or ± 1 % of reading, whichever is greater
Response time:
Typically, 9.5 minutes at 25 °C; changes with temperature
Sample pressure:
0.14 - 6 bar (to Preset Pressure Regulator)
Sample temperature:
5 - 50°C (41 - 122°F)
Storage conditions:
-20 - 60 °C
Weight:
20 kg without reagents and standards, 36.3 kg with reagents