Ambient temperature:
5 - 45 °C
Automatic cleaning:
Yes
Calibration:
Automatic with Self-Cleaning Technology
Communication: digital:
Modbus RTU, Modbus TCP/IP & Profibus (when the Profibus option is selected, the digital output signals are sent through the Profibus converter with its specific communication protocol)
Cycle Time:
Typically 7 minutes, 30 seconds
Data storage:
Previous 9,999 reaction data
Previous 99 fault events
Display:
High contrast 40 character x 16 line backlit LCD with LED backlight
EExp / Hazardous Location:
No
Exceedance tracking:
Full exceedance tracking up to 1,000 mg/L C
Humidity:
5 - 85 % (non-condensing)
Languages user interface:
English
Measurement method:
Infrared measurement of CO2 after oxidation
Measuring range:
250 mg/L C (ppm) with extended range capability 0 - 1000 ppm
Number of channels:
1
Number of sample streams:
Single process stream and manual grab sample
Number of streams:
Single stream
Options:
Grab Sample, Cleaning, Sample Sensor
Oxidation method:
Unique Two-Stage Advanced Oxidation Process (TSAO) using Hydroxyl Radicals
Parameter:
Measures TOC, TIC and TC
Particle size:
Up to 100 µm
Power requirements (Hz):
50 Hz
Power requirements (Voltage):
230 V AC
Range selection:
Automatic or manual range selection
Repeatability:
Standard range ±3% of reading or ±0.45 mg/L C, whichever is greater; 4% of reading or ±2 mg/L C (whichever is greater) for exceedance range
Sample inlet temperature:
2 - 60 °C
Service interval:
6 month service intervals
User interface:
Microcontroller with membrane keyboard
Warranty:
12 months
Weight:
46 kg
What's included?:
Includes: B3500e analyser, Tubing, Fuses, Ferrules, Drain, Acid & Base Dip Tubes, C02 Filter & B3500e User Manual