Accuracy:
± 0.02 pH
Cable connection:
Digital
Cable length:
10 m PUR (polyurethane) 4-conductor with one shield, rated to 105 °C
Calibration method:
Two point automatic, one point automatic, two point manual, one point manual
Communication:
Modbus
Compliance:
Hazardous location, Maritime, CE
Drift:
0.03 pH per 24 hours, non-cumulative
Electrode type:
General Purpose
Flow rate:
3 m per second, maximum
Housing material:
Stainless Steel
Immersion depth:
Submersible to 107 m/1050 kPa
Length:
271.3 mm
Measuring range:
-2.0 to 14.0 pH
Model:
pHD-S sc pH
Mounting:
Immersion
Operating temperature range:
-5 - 70 °C (23 - 158 °F) pHD and ORP
0 - 50 °C (32 - 122 °F) SS pHD
Before initial pH calibration, calibrate the temperature measurement when the sensor is in water or buffer which is at approximately the same temperature as the pH buffers (matches current recommendation)
Pressure range:
Max. 2 bar overpressure
Repeatability:
± 0.05 pH
Sensitivity:
± 0.01 pH
Sensor cable:
10 m (33 ft.) polyurethane, 4-conductor cable with one shield, rated to 105 °C (221°F)
Sensor thread:
1" NPT
Storage conditions:
0 - 95% relative humidity (non-condensing)
Temperature accuracy:
± 0.5 °C (± 0.9 °F)
Temperature compensation:
Automatic with NTC 300 Ω thermistor, or manually fixed at a user-entered temperature, additional selectable temperature correction factors (ammonia, morpholine, or user-defined pH/°C linear slope) available for pure water automatic compensation 0.0 - 50 °C
Temperature sensor:
NTC 300 Ω thermistor for automatic temperature compensation and analyzer temperature readout
Transmission distance:
100 m, maximum
1000 m, maximum when used with a termination box
Warranty:
24 months
Weight:
0.870 kg
Wetted Materials:
Stainless steel, PPS, glass, titanium, FKM/FPM o-ring