Accuracy:
With Standard Solution:
2% of the measured value + 1.0 mg/L
Altitude:
2000 m maximum
Cable length:
Standard: 2 m cable fixed from the edge of the enclosure
Extension cables are available as an option (data cable up to 50 m, power cable up to 5 m).
Certifications:
CE, UKCA, CMIM, FCC, ISED, certified to UL and CSA safety standards by TÜV
Controller compatibility:
Digital SC Controller. Controllers are sold separately.
Detection limit:
1 mg/L PO₄-P
Dimensions (H x W x D):
991 mm x 575 mm x 425 mm
Enclosure waterproof rating:
Rating: IP55, NEMA 3R UL50E
Material: PUR 66
Filter Configuration:
Use e.g. Filtrax
Flow rate:
0.5 - 20.0 L/h
Indoor/Outdoor:
Indoor or outdoor use
Measurement method:
Photometric (Ortho-phosphate ions react with vanadate-molybdate reagent and form a yellow dye.)
Measuring interval:
5, 10, 15, 20 or 30 minutes (user adjustable)
Measuring range:
1 - 75 mg/L PO₄-P
Number of channels:
2
Operating temperature range:
-20 - 45 °C; 95% relative humidity, non-condensing
Options:
Flow detection
Output:
Relay, analog outputs, network interface through SC Controller
Parameter:
Orthophospate
Permissible Chloride range:
≤ 1000 mg/L Cl-
pH Range:
5 - 9 pH
Pollution degree:
2
Power requirements (Voltage):
230 V AC
Power supply:
Mains power supplied by SC Controller or LQV155 power box.
Analyser and heated drain tubing: 230 VAC
Protection class:
Class I
Repeatability:
With Standard Solution:
2% of the measured value + 1.0 mg/L
Response time:
90% per measurement cycle for PO₄–P > 0.2 mg/L
80% per measurement cycle for PO₄–P ≤ 0.2 mg/L
Sample pressure:
2.5 bar (0.25 mPa) maximum
Sample quality:
Ultra filtrated or comparable
Sample temperature:
4 - 40 °C
Special remarks:
The liquid level in the basin must be below the bottom of the analyser.
Storage conditions:
-20 - 60 °C; 95% relative humidity, non-condensing
Warranty:
24 months
Weight:
Approximately 45 kg without chemicals
What's included?:
NP6000sc Phosphate Analyser, Mounting Hardware, User Manual