Accuracy:
Using standard solutions: 2 % + 0.015 mg/L
Cable length:
2 m from edge of enclosure
Channels:
1
Data types:
SC standard
Detection limit:
Using standard solutions: 0.015 mg/L
Dimensions (H x W x D):
720 mm x 540 mm x 370 mm
Enclosure waterproof rating:
IP54
Flow:
1.0 - 20.0 L/h
Housing type:
Indoor
Indoor/Outdoor:
Indoor
Material:
ASA/PC UV-resistant
Measurement method:
Two-beam photometer (yellow method)
Measuring interval:
10 - 120 min
Measuring range:
0.015 - 2 mg/L PO₄-P
Operating temperature range:
5 - 40 °C, 95 % relative humidity (non-condensing, non-corrosive)
Output:
Relay, analogue outputs, network interface through SC controller
Parameter:
Ortho-Phosphate
Permissible Chloride range:
Max Cl- concentration: 5000 mg/L
pH Range:
5 - 9 pH
Power requirements (Hz):
50/60 Hz
Power requirements (Voltage):
115 - 230 V AC, powered by Power Box or SC Controller
Power supply:
115 - 230 V AC, powered by Power Box or SC Controller
Pressure range:
-30 - 50 mbar with continous sample preparation; at overflow vessel
Reproducibility:
0.7 % + 0.005 mg/L
Response time:
10 min
Sample level:
Level of liquid in a basin with filtration probe must be below analyser
Sample quality:
Ultra filtrated or comparable
Sample temperature:
4 - 45 °C
Storage conditions:
-20 - 60 °C; 95 % relative humidity, non-condensing
Warranty:
24 months
Weight:
Without sample preparation system and without chemicals: 29 kg
What's included?:
Phosphax sc LR Phosphate Analyser, 3 month supply of all necessary consumables at 5 minute cycle time, user manuals