Acidity:
Less than 50 ppm (equivalent CaCO3)
Ambient temperature:
5 - 50 °C
Analogue outputs:
4 × (0 or 4) /20 mA [650 Ohms] / linear and dual / Smart
Communication capabilities:
Standard (4x 4-20 mA outputs)
Display:
Digital, 75 × 75 mm graphic with LED backlighting
European standards for EMC:
Conform EN 61326-1:2006, EN 61010-1:2001
Flow:
0.2–6 bar (3–87 psi), 5 L/h during sampling phase
Height (panel):
850 mm
Inlet:
Simple fittings for 6 mm O.D. tubing or 1/4" O.D. in PE-low density.
For 1/4" O.D. in PEHD-PTFE-SS, request imperial kit.
International standards:
U. L. (contact us for GOST Metrology)
Max. concentration of suspended solids in sample:
Below 10 ppm, no oil, no grease. For boiler sample type, install approx. 100 µm filter
Measuring range:
0.01 - 10000 ppb freely programmable
Mounting:
Wall mount enclosure
Number of channels:
1 Single channel
Number of relays:
2 programmable (concentration, temperature, minimum sample flow rate)
Outlet:
Barbed stem for 12 mm (1/2" I.D.) hose
Parameter:
Sodium
Power requirements (Hz):
50 - 60 Hz
Power requirements (Voltage):
100 - 240 V AC
Protection rating:
IP 65 (NEMA 4)
Repeatability with DIPA (within a 10° variation):
below 0.02 ppb or 1.5% reading, whichever is greater, within 10 °C variation
Response time:
180 s (t = 90%)
Sample temperature:
5 - 45°C (41 - 113°F)
Weight:
Empty canisters: Enclosure 23 kg
Full canisters: Enclosure 25 kg