Ambient temperature:
-20 - 60 °C
Analysis time:
Approx. 3 minutes
Detection limit:
TN: 20 µg/L
Dimensions:
440 mm x 380 mm x 700 mm (H x W x D)
Gas consumption:
200 mL/min
Gas specifications:
Oxygen or synthetic air: minimum 99.998% (4.8) at 3 - 10 bar
Measurement method:
Chemiluminescence
Measuring range:
TNb: 0 - 30000 mg/L
Model:
QP1680-TNb
Norms:
TNb: ASTM D8083, EN 12260, ISO 11905-2, NEN-ISO 20236
Number of sample positions:
96
Operating conditions:
20 - 30 °C; 20 - 80% relative humidity (non-condensing)
Oxidation method:
Catalytic combustion at 720 °C
Parameter:
TNb
Pollution degree:
2
Power:
Analyser: 750 W max.
PC: 90 W max.
Monitor: 100 W max.
Power supply:
Analyser: 100 - 240 VAC, 50/60 Hz, 16 A, with protective grounding
PC: 100 - 240 VAC, 50/60 Hz, 1.6 A, with protective grounding
Monitor: 100 - 240 VAC, 50/60 Hz, 1.6 A, with protective grounding
Repeatability:
Up to 10 mg/L TC, TIC, NPOC, TN: < 5%
> 10 mg/L TC, TIC, NPOC, TN: < 2%
Sample volume:
10 - 1000 µL
Temperature:
Furnace temperature max. 1050 °C (depending on configuration)
Warranty:
24 months
Weight:
Approx. 29 kg
What's included?:
TNb Instrument
Sample tray with 96 positions for 24 mL vials
Glass sample vials 24 mL (100 pcs)
USB 2.0 cable A-B
Set of tubing PTFE blue-stripe 1/8" OD length 2 m including nut & ferrule
Reagent container with screwcap for UPW, Acid and Waste
Power cord
USB storage device with software, test results and Full User Manual
Printed User Manual