Sigmatax 2 Control unit sampling & ultrasonic homogenisation

Sigmatax 2 Control unit sampling & ultrasonic homogenisation
Product #: LXV215
Unit Price Contact Hach

What's included?

1 set of operating instructions,1 maintenance calendar, 1 Factory Test Certificate

Specifications

Current output: 1 A 24 V, error signaling contact, floating
Delivery height: 7 m (for hose length 30 m, max 6 m)
Diameter: 184 mm probe
Dimensions (H x W x D): 560 mm x 366 mm x 212 mm
Enclosure waterproof rating: IP54
Hose length: 10 - 30 m (heated)
Interface: Service interface: RS 232
Length: 404 mm Sonde
Maintenance requirements: 0.5 h/week
Measuring interval: 12 to 20 min for up to 2 Phosphax sigma or TOCTAX process photometers
Measuring principle: Sample preperation only; Sample pumping controlled by pressure, homogenisation using ultrasound
Operating temperature range: 5 - 30 °C (Sample)
  5 - 40 °C control unit
Power consumption: 150 VA
Power requirements (Hz): 50 Hz
Power requirements (Voltage): 230 V AC
Service interval: 3 months
Warranty: 24 months
Weight: 12 kg control unit
  7.5kg probe with 10 m hose
What's included?: 1 set of operating instructions,1 maintenance calendar, 1 Factory Test Certificate