Accuracy:
Turbidity up to 1000 NTU:
without calibration < 5% of the measured value ±0.01 NTU
with calibration < 1% of the measured value ±0.01 NTU
Automatic wiper:
Yes
Body Material:
Stainless Steel
Cable length:
10 m (optional extension cables available)
Calibration method:
Turbidity: Formazin or Stablcal Standard (at 800 NTU).
Suspended Solids: Sample specific, based on gravimetric TSS analysis with a correction factor procedure.
Certifications:
CE
Controller:
Sensor only
Diameter:
60 mm
Flow:
Max. 3 m/s (the presence of air bubbles affects the measurement)
Length:
315 mm
Maintenance interval:
1 h/month
Material:
Optics carrier and sleeve: PVC black
Wiper arm: stainless steel 1.4581
Wiper rubber: silicone rubber (standard)
Optional: FKM/FPM (LZX578)
Wiper shaft: stainless steel 1.4571
Max. cable length Sensor-to-Controller:
100 m
Measuring principle:
Infrared Duo scattered light technique for color-independent
turbidity measurement Turbidity in accordance with
DIN EN ISO 7027 / TS equivalent DIN 38414
Measuring range:
TSS-content: 0.001 - 500 g/L; 0.001 - 500,000 mg/L (the upper range is dependent on the sample property)
Turbidity: 0.001 - 4000 NTU
Model:
Solitax highline sc
Mounting:
Insertion
Operating temperature range:
0 - 40 °C
Parameter:
High Range Suspended Solids, Turbidity
Pressure range:
Stainless steel insertion sensor: <6 bar or <60 m
Repeatability:
TSS content: < 3 %
Turbidity: < 1 %
Response time:
1 - 300 s adjustable
Special notes:
Installation note:
Distance sensor–wall
> 50 cm (Turbidity)
Warranty:
24 months
Weight:
Insertion stainless steel: 2.4 kg (5.29 lb.)
What's included?:
Turbidity/Suspended Solids probe, user manual