Accuracy:
Turbidity up to 1000 NTU:
without calibration < 5% of the measured value ±0.01 NTU
with calibration < 1% of the measured value ±0.01 NTU
Automatic wiper:
No
Body Material:
Stainless Steel
Cable length:
10 m (optional extension cables available)
Calibration method:
Turbidity Formazin or StablCal® Standard (at 800 NTU). Requires a calibration kit. Suspended Solids Sample specific: based on gravimetric TSS analysis with a correction factor procedure
Certifications:
CE
Controller:
Sensor only
Controller compatibility:
SC4500, SC1000, SC200
Diameter:
60 mm
Flow:
Max. 3 m/s (the presence of air bubbles affects the measurement)
Length:
315 mm
Maintenance interval:
1 h/month
Material:
stainless steel 1.4571
Max. cable length Sensor-to-Controller:
100 m
Measuring principle:
Infrared Duo scattered light technique for color-independent turbidity measurement Turbidity in accordance with DIN EN ISO 7027 / TS equivalent DIN 38414
Measuring range:
TSS content: 0.001 mg/L - 50 g/L
Turbidity: 0.001 - 4000 NTU
Model:
Solitax inline sc
Mounting:
Insertion
Operating temperature range:
0 - 40 °C
Parameter:
Suspended Solids, Turbidity
Pressure range:
Stainless steel insertion sensor: <6 bar or <60 m
Repeatability:
TSS content: < 3%
Turbidity: < 1%
Response time:
1 - 300 s adjustable
Special notes:
Installation note:
Distance sensor–wall
> 10cm
Installation note:
Distance sensor–wall
> 50 cm (Turbidity)
Warranty:
24 months
Weight:
Insertion stainless steel: 2.4 kg (5.29 lb.)
What's included?:
Turbidity/Suspended Solids probe, user manual