Accuracy:
Turbidity up to 1000 FNU/NTU: < 5 % of measurement value or ±0.01 NTU, whichever is greater
Ambient temperature:
0 - 50 °C
Application:
Hazardous locations
Automatic wiper:
No
Cable length:
10 m
Calibration:
Turbidity: Factory calibrated
Suspended solids: To be calibrated by customer on site
Zero point: Permanently calibrated in the factory
Calibration method:
Turbidity: Formazin or Stablcal Standard (at 800 NTU). Requires a calibration kit.
Suspended Solids: Sample specific, based on gravimetric analysis with a correction factor procedure.
Controller compatibility:
SC200, SC1000, SC4500. All controllers sold separately
Diameter:
48.5 mm
Flow:
Max. 3 m/s (the presence of air bubbles affects the measurement)
Installation style:
Tank Immersion
Length:
385 mm
Maintenance interval:
1 h/month
Material:
stainless steel 1.4571
Max Temperature:
50 °C
Measurement method:
Combined multiple beam alternating light method with infrared diode system and beam focusing
Turbidity (TRB): 2-channel 90º scattered light measurement in accordance with DIN EN ISO 7027, wavelength = 860 nm
Solids (TS): 90º and 120º scattered light measurement, wavelength = 860 nm
Measuring principle:
Combined multiple beam alternating
light method with infrared diode
system and beam focussing
Turbidity (TRB): 2-channel 90° scattered light measurement
in accordance with DIN EN ISO 7027,
wavelength = 860 nm
Solids (TS): 120° scattered light measurement,
wavelength = 860 nm
Measuring range:
Turbidity (TRB): 0.001 - 9999 NTU
Solids (TSS): 0.001 - 500 g/L
Model:
TSS EX1 sc
Mounting configurations:
Immersion
Operating temperature range:
-10 - 50 °C
Parameter:
Turbidity, Suspended Solids
Pressure range:
10 bar or 100 m
Repeatability:
TSS content: < 4 %
Turbidity: < 3 %
Response time:
1 - 300 s adjustable
Special notes:
Installation note:
Distance sensor–wall
> 50 cm (Turbidity)
> 10 cm (TSS)
Warranty:
24 months
Weight:
approx. 2.7 kg
What's included?:
Turbidity & Suspended Solids sensor, user manual