TSS HT sc Triclamp Suspended solids probe, stainless steel, for high temperatures, ball valve fitting

TSS HT sc Triclamp Suspended solids probe, stainless steel, for high temperatures, ball valve fitting
Product #: LXV325.99.20001
Unit Price Contact Hach

TSS sc probes have been specially developed for industrial applications

Measures both turbidity and suspended solids

8 measurement signals cover the total range and deliver measured values in conformity with the relevant standards

Excellent stability thanks to comprehensive compensation for interference factors

TSS sc has a unique compensation system to overcome the effects of air bubbles

What's included?

Turbidity & Suspended Solids sensor, user manual

Specifications

Accuracy: Turbidity up to 1000 FNU/NTU: < 5% of the measured value ±0.01 NTU
Ambient temperature: 0 - 60 °C (briefly 80 °C)
Application: High temperature
Automatic wiper: No
Cable length: 10 m (optional extension cables available)
Calibration: Turbidity (TRB): Factory calibrated
Solids (TS): To be calibrated by customer on site
Zero point: Permanently calibrated in the factory
Calibration method: Turbidity: Formazin or Stablcal Standard (at 800 NTU). Requires a calibration kit.
Suspended Solids: Sample specific, based on gravimetric analysis with a correction factor procedure.
Controller compatibility: SC200, SC1000, SC4500. All controllers sold separately
Diameter: 40 mm
Flow: Max. 3 m/s (the presence of air bubbles affects the measurement)
Installation style: Weld On TriClamp Fitting, High Temperature
Length: Inline sensor (TriClamp): 332 mm
Maintenance interval: 1 h/month
Material: Gasket: FKM; Wiper: PA (GF), TPV
Max Temperature: 90 °C
Measurement method: Combined multiple beam alternating light method with infrared diode system and beam focusing
Turbidity (TRB): 2-channel 90º scattered light measurement in accordance with DIN EN ISO 7027, wavelength = 860 nm
Solids (TS): 90º and 120º scattered light measurement, wavelength = 860 nm
Measuring principle: Combined multiple beam alternating
light method with infrared diode
system and beam focussing
Turbidity (TRB): 2-channel 90° scattered light measurement
in accordance with DIN/EN 27027/ISO 7027,
wavelength = 860 nm
Solids (TS): 120° scattered light measurement,
wavelength = 860 nm
Measuring range: Turbidity (TRB): 0.001 - 9999 NTU
Solids (TSS): 0.001 - 500 g/L
Model: TSS HT sc
Mounting configurations: TriClamp
Operating temperature range: 0 - 90 °C
Parameter: Turbidity, Suspended Solids
Pressure range: < 10 bar or < 100 m
Repeatability: TSS content: < 4 %
Turbidity: < 3 %
Response time: 1 - 300 s adjustable
Special Features: with Triclamp
Special notes: Installation note:
Distance sensor–wall
> 50 cm (Turbidity)
> 10 cm (TSS)
Warranty: 24 months
Weight: approx. 1.6 kg
What's included?: Turbidity & Suspended Solids sensor, user manual