The AS950 All-Weather Refrigerated Sampler is the perfect solution for easy and reliable sample collection in all conditions. With its specially designed air-sensing thermostat, it preserves samples regardless of outside temperatures and conditions, ensuring that your samples are safe and accurate.
In addition to its advanced temperature control, the AS950 All-Weather Refrigerated Sampler features a large, full-color display, intuitive single-screen programming, and USB data transfer capabilities that simplify and streamline the entire sampling process. You can even copy programs from sampler to sampler, eliminating tedious programming and saving time.
Investing in the AS950 All-Weather Refrigerated Sampler gives you the easiest and most intuitive operation, the most convenient data transfer and programming, and the confidence in your sampling process you need, no matter what the weather. Order yours today and simplify your sampling process.
The AS950 All-Weather Refrigerated Sampler is the perfect solution for easy and reliable sample collection in all conditions. With its specially designed air-sensing thermostat, it preserves samples regardless of outside temperatures and conditions, ensuring that your samples are safe and accurate.
In addition to its advanced temperature control, the AS950 All-Weather Refrigerated Sampler features a large, full-color display, intuitive single-screen programming, and USB data transfer capabilities that simplify and streamline the entire sampling process. You can even copy programs from sampler to sampler, eliminating tedious programming and saving time.
Investing in the AS950 All-Weather Refrigerated Sampler gives you the easiest and most intuitive operation, the most convenient data transfer and programming, and the confidence in your sampling process you need, no matter what the weather. Order yours today and simplify your sampling process.
Base Type
With controller compartment heater
Power Requirements (Voltage)
230 V
Sample container
1 x 20 L
Can't find the combination you need?
Explore the AS950 All Weather Samplers family to find the right product for your requirements.
The AS950 All-Weather Refrigerated Sampler is the perfect solution for easy and reliable sample collection in all conditions. With its specially designed air-sensing thermostat, it preserves samples regardless of outside temperatures and conditions, ensuring that your samples are safe and accurate.
In addition to its advanced temperature control, the AS950 All-Weather Refrigerated Sampler features a large, full-color display, intuitive single-screen programming, and USB data transfer capabilities that simplify and streamline the entire sampling process. You can even copy programs from sampler to sampler, eliminating tedious programming and saving time.
Investing in the AS950 All-Weather Refrigerated Sampler gives you the easiest and most intuitive operation, the most convenient data transfer and programming, and the confidence in your sampling process you need, no matter what the weather. Order yours today and simplify your sampling process.
Easiest and most intuitive operation
The large full colour display and intuitive programming give you access to all your programmable criteria on a single screen - eliminating scrolling through menus and supporting error-free operation.
Most convenient data transfer and programming available
The AS950 is the only sampler that utilises a USB drive to upload and download data and copy programs from one sampler to another.
Confidence in your sampling process
The program status screen instantly communicates alarms, missed samples and program progress for quick and easy troubleshooting.
Resists corrosion
The All-Weather Refrigerated (AWR) sampler base is designed to endure humid and highly corrosive environments, minimising damage caused by corrosive gases, rodents, and standing water to guarantee environmental integrity.
Applications
Discharge
Flow Monitoring
Groundwater
Industrial Wastewater
Specifications
Communication capabilitiesUSB and optional RS485 (Modbus)
Display1/4 VGA, Color; self-prompting/menu-driven program
Housing materialPC/ABS blend, NEMA 6, IP68, corrosion and ice resistant (Controller)
Low-density Polyethylene with UV inhibitor (Sampler)
Pump flow rate4.8 L/min at 1 m vertical lift with 3/8" intake tube typical
The AS950 All-Weather Refrigerated Sampler is the perfect solution for easy and reliable sample collection in all conditions. With its specially designed air-sensing thermostat, it preserves samples regardless of outside temperatures and conditions, ensuring that your samples are safe and accurate.
In addition to its advanced temperature control, the AS950 All-Weather Refrigerated Sampler features a large, full-color display, intuitive single-screen programming, and USB data transfer capabilities that simplify and streamline the entire sampling process. You can even copy programs from sampler to sampler, eliminating tedious programming and saving time.
Investing in the AS950 All-Weather Refrigerated Sampler gives you the easiest and most intuitive operation, the most convenient data transfer and programming, and the confidence in your sampling process you need, no matter what the weather. Order yours today and simplify your sampling process.
Easiest and most intuitive operation
The large full colour display and intuitive programming give you access to all your programmable criteria on a single screen - eliminating scrolling through menus and supporting error-free operation.
Most convenient data transfer and programming available
The AS950 is the only sampler that utilises a USB drive to upload and download data and copy programs from one sampler to another.
Confidence in your sampling process
The program status screen instantly communicates alarms, missed samples and program progress for quick and easy troubleshooting.
Resists corrosion
The All-Weather Refrigerated (AWR) sampler base is designed to endure humid and highly corrosive environments, minimising damage caused by corrosive gases, rodents, and standing water to guarantee environmental integrity.
Applications
Discharge
Flow Monitoring
Groundwater
Industrial Wastewater
Specifications
Communication capabilitiesUSB and optional RS485 (Modbus)
Display1/4 VGA, Color; self-prompting/menu-driven program
Housing materialPC/ABS blend, NEMA 6, IP68, corrosion and ice resistant (Controller)
Low-density Polyethylene with UV inhibitor (Sampler)
Pump flow rate4.8 L/min at 1 m vertical lift with 3/8" intake tube typical
The custom-designed air-sensing thermostat controls temperature in accordance with USEPA and international guidelines, preserving samples regardless of outside temperatures and conditions.
Influent for Wastewater
Municipal Water
Pre-Treatment
Primary Treatment
Production Water
QA/QC Lab
Regulations
Secondary Treatment
Source Water
Surface Water
Storm Water
Water Cycle
Water Reuse
Suction Hose Outer Diameter
1/4" and 3/8" ID PTFE-lined Polyethylene with protective outer cover available in clear and black.
Weight86 kg
Data loggingSAMPLE HISTORY - Stores up to 4000 entries for sample time stamp, bottle number and sample status (success, bottle full, rinse error, user abort, distributor error, pump fault, purge fail, sample timeout, power fail and low main battery).
MEASUREMENTS - Stores up to 325,000 entries for selected measurement channels in accordance with the selected logging interval.
EVENTS - Ability to store up to 2000 entries in Sample History logging. Records Power On, Power Fail, Firmware Updated, Pump Fault, Distributor Arm Error, Low Memory Battery, Low Main Battery, User On, User Off, Program Started, Program Resumed, Program Halted, Program Completed, Grab Sample, Tube Change Required, Sensor Communication Errors, Cooling Failed, Heating Failed, Thermal Error Corrected.
Refrigeration components and plumbingCorrosion protected with conformal all exposed copper tubing is insulated to avoid sweating and condensation
Sample intervalSelectable in single increments from 1 to 9999 flow pulses or 1 to 9999 minutes in one minute increments
Manual samplingInitiates a sample collection independent of program in progress
Sample pumpHigh-speed peristaltic, dual roller, with 0.95 ID x 1.6 OD cm (3/8" ID x 5/8" OD) pump tube; Pump body IP37, Polycarbonate cover
Sample volume repeatabilityTypical: ±5% of 200 mL sample volume with: 4.6 m vertical lift, 4.9 m of 3/8" Vinyl intake tube, single bottle, full bottle shut-off at room temperature and 1524 m elevation
Operating modeContinuous or non-continuous
Rain Gauge InputNo
IntakeStrainers: Choice of PTFE and 316 stainless steel construction, or all 316 stainless steel in standard size, high velocity, and low profile for shallow depth applications
Purge: Air purged automatically before and after each sample; duration automatically compensates for varying intake line lengths
Rinse: Intake line automatically rinsed with source liquid prior to each sample, from 1 to 3 rinses
Retries or Fault: Sample collection cycle automatically repeated from 1 to 3 times if sample not obtained on initial attempt
CertificationsCE, UL
CE, UL, CSA
Sampling modePacing: Time weighted, Flow weighted, Time table, Flow table, Event
Distribution: Single bottle composite, multi-bottle composite, multi-bottle discrete, bottles per sample, samples per bottle or a combination of bottles per sample and samples per bottle
MemorySample history: 4000 records; Data log: 325,000 records; Event log: 2000 records
Sample triggerWhen equipped with flow sensor or pH/temperature sensor or peripheral monitoring options, sampling can be triggered upon an upset condition when field selectable limits are exceeded
Cooling systemTop mounted compressor and fan-forced air cooled condenser; 11/5 HP 115 VAC: 115 °C thermal overload protector, 7.1 locked rotor amps; 230 VAC: 120 °C thermal overload protector, 7.6 A peak start current; 3-sided wraparound plate type evaporator; Rigid foam insulation: 7.6 cm sides, 12.7 cm top, 15.2 cm bottom; Lockable lid to prevent tampering with programming; Recovery Time: Sampler temperature recovers to 4 °C within 5 minutes after door has been held open for one minute in 24 °C ambient environment while in an active cooling cycle; Pull Down Time: from 24 °C to 4 °C , 20 minutes; Temperature Control: 4 °C (±0.8 °C)
Power requirements (Hz)50 - 60 Hz
What's included?1 ea. AS950 controller on 230 V All Weather base with controller compartment heater 1 ea. Bottle kit (AW010060), includes 1 x 20 L PE bottle w/cap (6494), composite tube support (8838) and full bottle shut-off (8847) 1 ea. 7 m vinyl intake tubing (920) 1 ea. strainer (926)
Power Requirements (Voltage)230 V
Special FeaturesWith Heater and Lock
TubingPump tubing: 9.5 mm ID x 15.9 mm OD Silicone
Intake tubing: 1.0 - 4.75 m minimum length, 1/4" or 3/8" ID Vinyl or 3/8" ID PTFE-lined Polyethylene with protective outer cover (black or clear)
Sensor Ports0
Automatic shutdownMultiple Bottle Mode: After complete revolution of distributor arm (unless Continuous Mode is selected) Composite Mode: After preset number of samples have been delivered to composite container, from 1 to 999 samples, or upon full container.
AlarmConfigurable alarms that show on status screen and are recorded in diagnostics alarm logs. Alarms can be set for system diagnostics and logging such as program end, sample complete, missed samples and full bottle. Channel alarms are setpoint alarms for the recorded measurements (channels), such as pH, level and power supply voltage.
Bottle SetComposite 20 L PE Bottle
Base Configuration230V base w/heater and lock, EU
Languages user interfaceEnglish, French, Spanish, Italian, German, Portuguese, Chinese
Sample container1 x 20 L
Sample volume accuracyTypical: ±5% of 200 mL sample volume with: 4.6 m vertical lift, 4.9 m of 3/8" Vinyl intake tube, single bottle, full bottle shut-off at room temperature and 1524 m elevation
TemperatureOperating: AWRS: 0 - 50 °C AWRS with optional controller compartment heater: -40 - 50 °C AWRS with controller compartment heater and AC battery back up: -15 - 40 °C Storage: -30 - 60 °C
User interfaceChinese, English, French, German, Italian, Spanish, Portuguese
Membrane switch keypad with 2 multiple function soft keys
Sample volumeProgrammable in 10 mL increments from 10 - 10,000 mL
Power Supply230 V
DiagnosticsView event and alarm logs as well as maintenance diagnostics
Probes Included?No
Warranty24 months
InputsOne 0/4-20 mA input for flow pacing
Vertical lift8.5 m using 8.8 m maximum of 3/8" Vinyl intake tube at sea level at 20 to 25 °C
Internal clock±1 second per day at 25 °C
Rain Gauge EnabledNo
Base TypeWith controller compartment heater
SensorsNone
ProgramsDual Programs: Up to 2 sample programs can be run sequentially, in parallel, or according to day of week scheduling; enabling a single sampler to function like multiple samplers
Status messagesCommunicates what program is running, if there are any missed samples, when the next sample will be taken, how many samples remain, number of logged channels, time of last measurement, memory available, number of active channels, if alarms were triggered, when alarms were triggered, active sensors and cabinet temperature
Sample transport velocity0.9 m/s with 4.6 m vertical lift, 4.9 m of 3/8" Vinyl intake tubing, 21 °C and 1524 m elevation
Dimensions (H x W x D)1300 mm x 760 mm x 810 mm
Factory Installed OptionsTWO SENSOR PORTS Accepts Hach digital Differential pH, Hach digital AV9000 analyser with submerged area velocity flow and/or Hach digital US9000 ultrasonic level sensors , RAIN/RS485 PORT Accepts Hach Rain Gauge (not included) or can be used as RS485 communications , NON-CONTACT LIQUID DETECT Sample volume accuracy for applications that require complete tubing replacement , Please contact your local Hach representative for details.
Dimensions76 cm x 81 cm x 130 cm
Accurate and consistent sample preservations
The custom-designed air-sensing thermostat controls temperature in accordance with USEPA and international guidelines, preserving samples regardless of outside temperatures and conditions.
Influent for Wastewater
Municipal Water
Pre-Treatment
Primary Treatment
Production Water
QA/QC Lab
Regulations
Secondary Treatment
Source Water
Surface Water
Storm Water
Water Cycle
Water Reuse
Suction Hose Outer Diameter
1/4" and 3/8" ID PTFE-lined Polyethylene with protective outer cover available in clear and black.
Weight86 kg
Data loggingSAMPLE HISTORY - Stores up to 4000 entries for sample time stamp, bottle number and sample status (success, bottle full, rinse error, user abort, distributor error, pump fault, purge fail, sample timeout, power fail and low main battery).
MEASUREMENTS - Stores up to 325,000 entries for selected measurement channels in accordance with the selected logging interval.
EVENTS - Ability to store up to 2000 entries in Sample History logging. Records Power On, Power Fail, Firmware Updated, Pump Fault, Distributor Arm Error, Low Memory Battery, Low Main Battery, User On, User Off, Program Started, Program Resumed, Program Halted, Program Completed, Grab Sample, Tube Change Required, Sensor Communication Errors, Cooling Failed, Heating Failed, Thermal Error Corrected.
Refrigeration components and plumbingCorrosion protected with conformal all exposed copper tubing is insulated to avoid sweating and condensation
Sample intervalSelectable in single increments from 1 to 9999 flow pulses or 1 to 9999 minutes in one minute increments
Manual samplingInitiates a sample collection independent of program in progress
Sample pumpHigh-speed peristaltic, dual roller, with 0.95 ID x 1.6 OD cm (3/8" ID x 5/8" OD) pump tube; Pump body IP37, Polycarbonate cover
Sample volume repeatabilityTypical: ±5% of 200 mL sample volume with: 4.6 m vertical lift, 4.9 m of 3/8" Vinyl intake tube, single bottle, full bottle shut-off at room temperature and 1524 m elevation
Operating modeContinuous or non-continuous
Rain Gauge InputNo
IntakeStrainers: Choice of PTFE and 316 stainless steel construction, or all 316 stainless steel in standard size, high velocity, and low profile for shallow depth applications
Purge: Air purged automatically before and after each sample; duration automatically compensates for varying intake line lengths
Rinse: Intake line automatically rinsed with source liquid prior to each sample, from 1 to 3 rinses
Retries or Fault: Sample collection cycle automatically repeated from 1 to 3 times if sample not obtained on initial attempt
CertificationsCE, UL
CE, UL, CSA
Sampling modePacing: Time weighted, Flow weighted, Time table, Flow table, Event
Distribution: Single bottle composite, multi-bottle composite, multi-bottle discrete, bottles per sample, samples per bottle or a combination of bottles per sample and samples per bottle
MemorySample history: 4000 records; Data log: 325,000 records; Event log: 2000 records
Sample triggerWhen equipped with flow sensor or pH/temperature sensor or peripheral monitoring options, sampling can be triggered upon an upset condition when field selectable limits are exceeded
Cooling systemTop mounted compressor and fan-forced air cooled condenser; 11/5 HP 115 VAC: 115 °C thermal overload protector, 7.1 locked rotor amps; 230 VAC: 120 °C thermal overload protector, 7.6 A peak start current; 3-sided wraparound plate type evaporator; Rigid foam insulation: 7.6 cm sides, 12.7 cm top, 15.2 cm bottom; Lockable lid to prevent tampering with programming; Recovery Time: Sampler temperature recovers to 4 °C within 5 minutes after door has been held open for one minute in 24 °C ambient environment while in an active cooling cycle; Pull Down Time: from 24 °C to 4 °C , 20 minutes; Temperature Control: 4 °C (±0.8 °C)
Power requirements (Hz)50 - 60 Hz
What's included?1 ea. AS950 controller on 230 V All Weather base with controller compartment heater 1 ea. Bottle kit (AW010060), includes 1 x 20 L PE bottle w/cap (6494), composite tube support (8838) and full bottle shut-off (8847) 1 ea. 7 m vinyl intake tubing (920) 1 ea. strainer (926)
Power Requirements (Voltage)230 V
Special FeaturesWith Heater and Lock
TubingPump tubing: 9.5 mm ID x 15.9 mm OD Silicone
Intake tubing: 1.0 - 4.75 m minimum length, 1/4" or 3/8" ID Vinyl or 3/8" ID PTFE-lined Polyethylene with protective outer cover (black or clear)
Sensor Ports0
Automatic shutdownMultiple Bottle Mode: After complete revolution of distributor arm (unless Continuous Mode is selected) Composite Mode: After preset number of samples have been delivered to composite container, from 1 to 999 samples, or upon full container.
AlarmConfigurable alarms that show on status screen and are recorded in diagnostics alarm logs. Alarms can be set for system diagnostics and logging such as program end, sample complete, missed samples and full bottle. Channel alarms are setpoint alarms for the recorded measurements (channels), such as pH, level and power supply voltage.
Bottle SetComposite 20 L PE Bottle
Base Configuration230V base w/heater and lock, EU
Languages user interfaceEnglish, French, Spanish, Italian, German, Portuguese, Chinese
Sample container1 x 20 L
Sample volume accuracyTypical: ±5% of 200 mL sample volume with: 4.6 m vertical lift, 4.9 m of 3/8" Vinyl intake tube, single bottle, full bottle shut-off at room temperature and 1524 m elevation
TemperatureOperating: AWRS: 0 - 50 °C AWRS with optional controller compartment heater: -40 - 50 °C AWRS with controller compartment heater and AC battery back up: -15 - 40 °C Storage: -30 - 60 °C
User interfaceChinese, English, French, German, Italian, Spanish, Portuguese
Membrane switch keypad with 2 multiple function soft keys
Sample volumeProgrammable in 10 mL increments from 10 - 10,000 mL
Power Supply230 V
DiagnosticsView event and alarm logs as well as maintenance diagnostics
Probes Included?No
Warranty24 months
InputsOne 0/4-20 mA input for flow pacing
Vertical lift8.5 m using 8.8 m maximum of 3/8" Vinyl intake tube at sea level at 20 to 25 °C
Internal clock±1 second per day at 25 °C
Rain Gauge EnabledNo
Base TypeWith controller compartment heater
SensorsNone
ProgramsDual Programs: Up to 2 sample programs can be run sequentially, in parallel, or according to day of week scheduling; enabling a single sampler to function like multiple samplers
Status messagesCommunicates what program is running, if there are any missed samples, when the next sample will be taken, how many samples remain, number of logged channels, time of last measurement, memory available, number of active channels, if alarms were triggered, when alarms were triggered, active sensors and cabinet temperature
Sample transport velocity0.9 m/s with 4.6 m vertical lift, 4.9 m of 3/8" Vinyl intake tubing, 21 °C and 1524 m elevation
Dimensions (H x W x D)1300 mm x 760 mm x 810 mm
Factory Installed OptionsTWO SENSOR PORTS Accepts Hach digital Differential pH, Hach digital AV9000 analyser with submerged area velocity flow and/or Hach digital US9000 ultrasonic level sensors , RAIN/RS485 PORT Accepts Hach Rain Gauge (not included) or can be used as RS485 communications , NON-CONTACT LIQUID DETECT Sample volume accuracy for applications that require complete tubing replacement , Please contact your local Hach representative for details.