EZ3508sc Fluoride Ion-Selective Electrode (ISE) Standard Addition Analyser, 1 stream, 4x mA
Product Number: EZ3508.9700140T
Fluoride measurement in wastewater and industrial process water can be distorted by matrix effects from ions such as calcium and aluminum. The EZ3508sc uses a standard addition method to compensate for these interferences, delivering accurate, repeatable fluoride results while minimising reagent consumption.
Designed for reliable operation, gel-filled ISE electrodes eliminate electrolyte refilling and automated cleaning and calibration reduce hands-on maintenance. Digital communication protocols integrate the analyser into your network for immediate access to results, diagnostics and status.
Benefits: - Standard addition compensates for matrix effects in complex samples - Accurate, repeatable fluoride results with low reagent consumption - Gel-filled ISE electrodes eliminate electrolyte refilling - Automated cleaning and calibration reduce manual service - Digital connectivity for fast access to results, diagnostics and status
Fluoride measurement in wastewater and industrial process water can be distorted by matrix effects from ions such as calcium and aluminum. The EZ3508sc uses a standard addition method to compensate for these interferences, delivering accurate, repeatable fluoride results while minimising reagent consumption.
Designed for reliable operation, gel-filled ISE electrodes eliminate electrolyte refilling and automated cleaning and calibration reduce hands-on maintenance. Digital communication protocols integrate the analyser into your network for immediate access to results, diagnostics and status.
Benefits: - Standard addition compensates for matrix effects in complex samples - Accurate, repeatable fluoride results with low reagent consumption - Gel-filled ISE electrodes eliminate electrolyte refilling - Automated cleaning and calibration reduce manual service - Digital connectivity for fast access to results, diagnostics and status
APPAZ0080018Moduplex, 8 Streams, Pinch Valve, 7/9" OD
Overview
Fluoride measurement in wastewater and industrial process water can be distorted by matrix effects from ions such as calcium and aluminum. The EZ3508sc uses a standard addition method to compensate for these interferences, delivering accurate, repeatable fluoride results while minimising reagent consumption.
Designed for reliable operation, gel-filled ISE electrodes eliminate electrolyte refilling and automated cleaning and calibration reduce hands-on maintenance. Digital communication protocols integrate the analyser into your network for immediate access to results, diagnostics and status.
Benefits: - Standard addition compensates for matrix effects in complex samples - Accurate, repeatable fluoride results with low reagent consumption - Gel-filled ISE electrodes eliminate electrolyte refilling - Automated cleaning and calibration reduce manual service - Digital connectivity for fast access to results, diagnostics and status
Consistent Results in Challenging Conditions
Fluctuating water quality often compromises conventional monitoring, leading to unstable readings. The EZ3508sc uses discontinuous and standard addition modes to compensate for these matrix interferences. You get accurate, continuous fluoride data for stable process control, even as conditions change.
Reduced Maintenance and Manual Effort
Gel-filled ion selective electrodes eliminate electrolyte refilling, while automated cleaning and calibration reduce routine service needs. Discontinuous standard addition method optimises buffer and reagent consumption; lowering site visits, operating costs, and overall maintenance effort.
Seamless Integration and Connectivity
Isolated data makes process control and reporting difficult. The integrated controller platform supports standard digital communication protocols for easy connection to your existing network. You gain immediate access to critical insights, supporting faster decision-making and remote troubleshooting.
Streamlined Operation and Usability
Complex instrumentation often leads to training bottlenecks and operational errors. The unified controller interface simplifies setup and daily interaction. Operators of any experience level can easily manage the system, reducing training requirements and preventing errors during routine checks.
Applications
Municipal Water
Specifications
What's included?EZ3508sc Fluoride Ion-Selective Electrode (ISE) Standard Addition Analyser, 1 stream, 4x mA, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode, 1 x empty 5L Reagent Container with Fittings (for F Standard Solution) and 1 x empty 10L Reagent Container with Fittings (for Buffer Solution)
Number of sample streamsStandard: 1 Optional accessory for 2, 4, or 8 sample streams
Demineralised waterInternal dilution option: Rinse water and Dilution water
Measurement methodIon-Selective Electrode (standard addition)
Discontinuous measurement by combined Ion-Selective Electrode with standard addition
Documents & Downloads
Data Sheets
Data Sheet
EZ3000sc/EZ3500sc Series Analysers Service Data Sheet
Fluoride measurement in wastewater and industrial process water can be distorted by matrix effects from ions such as calcium and aluminum. The EZ3508sc uses a standard addition method to compensate for these interferences, delivering accurate, repeatable fluoride results while minimising reagent consumption.
Designed for reliable operation, gel-filled ISE electrodes eliminate electrolyte refilling and automated cleaning and calibration reduce hands-on maintenance. Digital communication protocols integrate the analyser into your network for immediate access to results, diagnostics and status.
Benefits: - Standard addition compensates for matrix effects in complex samples - Accurate, repeatable fluoride results with low reagent consumption - Gel-filled ISE electrodes eliminate electrolyte refilling - Automated cleaning and calibration reduce manual service - Digital connectivity for fast access to results, diagnostics and status
Consistent Results in Challenging Conditions
Fluctuating water quality often compromises conventional monitoring, leading to unstable readings. The EZ3508sc uses discontinuous and standard addition modes to compensate for these matrix interferences. You get accurate, continuous fluoride data for stable process control, even as conditions change.
Reduced Maintenance and Manual Effort
Gel-filled ion selective electrodes eliminate electrolyte refilling, while automated cleaning and calibration reduce routine service needs. Discontinuous standard addition method optimises buffer and reagent consumption; lowering site visits, operating costs, and overall maintenance effort.
Seamless Integration and Connectivity
Isolated data makes process control and reporting difficult. The integrated controller platform supports standard digital communication protocols for easy connection to your existing network. You gain immediate access to critical insights, supporting faster decision-making and remote troubleshooting.
Streamlined Operation and Usability
Complex instrumentation often leads to training bottlenecks and operational errors. The unified controller interface simplifies setup and daily interaction. Operators of any experience level can easily manage the system, reducing training requirements and preventing errors during routine checks.
Applications
Municipal Water
Specifications
What's included?EZ3508sc Fluoride Ion-Selective Electrode (ISE) Standard Addition Analyser, 1 stream, 4x mA, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode, 1 x empty 5L Reagent Container with Fittings (for F Standard Solution) and 1 x empty 10L Reagent Container with Fittings (for Buffer Solution)
Number of sample streamsStandard: 1 Optional accessory for 2, 4, or 8 sample streams
Demineralised waterInternal dilution option: Rinse water and Dilution water
Measurement methodIon-Selective Electrode (standard addition)
Discontinuous measurement by combined Ion-Selective Electrode with standard addition
Documents & Downloads
Data Sheets
Data Sheet
EZ3000sc/EZ3500sc Series Analysers Service Data Sheet
Instrument airDry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
CertificationsCE, ETL certified to UL and CSA safety standards, UKCA
InterferencesMetal ions like aluminium [(Al)³+] > 72 mg/L, calcium [(Ca)²+] > 108 mg/L and iron [(Fe)²+/(Fe)³+] > 150 mg/L. Fats, oil, proteins, surfactants and tar.
CalibrationAutomatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced
Dimensions (H x W x D)688 mm x 460 mm x 340 mm
Warranty24 months
Power100 - 240 VAC, 50/60 Hz Max. power consumption: 120 VA
DrainAtmospheric pressure, vented, min. Ø 32 mm
Analogue outputsActive 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard 4, optional: 8
Reagent RequirementsKeep between 10 - 30 °C (50 - 86 °F)
WeightMax. 35 kg (77 lb)
Ambient temperature10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Digital outputsRelays: 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150)
Ethernet Connections: Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP
Instrument airDry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
CertificationsCE, ETL certified to UL and CSA safety standards, UKCA
InterferencesMetal ions like aluminium [(Al)³+] > 72 mg/L, calcium [(Ca)²+] > 108 mg/L and iron [(Fe)²+/(Fe)³+] > 150 mg/L. Fats, oil, proteins, surfactants and tar.
CalibrationAutomatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced