EZ4000 Series Analysers are single-parameter titrators that achieve excellent precision and accuracy. Depending on the parameter of interest, the measuring range and the water matrix the EZ4000 Series may use either potentiometry i.e. acid-base, redox or precipitation titration, where the endpoint is determined by a change in a specific variable, or photocolorimetry, where colour change is used to determine the endpoint of the titration. The high precision dispenser for addition, robust peristaltic pumps for sampling and drain, and carefully designed liquid pathways all add up to the highest performance for industrial and environmental analysis needs.
Results you can rely on Smart automatic features for validation, priming and cleaning are embedded in the controller software and contribute to analytical performance, maximised uptime and negligible operator invervention. Sample lines and analysis vessel are cleaned with demineralised water to eliminate cross contamination between samples. Electronic and wet-chemical part of the analyser are strictly separated. A transparent door allows for instant visual inspection of the wet part.
Flexibility that meets your needs EZ Series Chloride Analysers come in an attractive, ergonomic mainframe with a compact footprint. All hardware is controlled by the integrated industrial panel PC. The modular build allows for the analyser to match your application and operational needs.
• The standard measuring range can be narrowed by a different calibration range or extended via internal dilution options. • Analogue and digital output options • Multiple stream analysis for up to 8 sample streams
There are many additional options available. Please contact Hach for more details.
EZ4000 Series Analysers are single-parameter titrators that achieve excellent precision and accuracy. Depending on the parameter of interest, the measuring range and the water matrix the EZ4000 Series may use either potentiometry i.e. acid-base, redox or precipitation titration, where the endpoint is determined by a change in a specific variable, or photocolorimetry, where colour change is used to determine the endpoint of the titration. The high precision dispenser for addition, robust peristaltic pumps for sampling and drain, and carefully designed liquid pathways all add up to the highest performance for industrial and environmental analysis needs.
Results you can rely on Smart automatic features for validation, priming and cleaning are embedded in the controller software and contribute to analytical performance, maximised uptime and negligible operator invervention. Sample lines and analysis vessel are cleaned with demineralised water to eliminate cross contamination between samples. Electronic and wet-chemical part of the analyser are strictly separated. A transparent door allows for instant visual inspection of the wet part.
Flexibility that meets your needs EZ Series Chloride Analysers come in an attractive, ergonomic mainframe with a compact footprint. All hardware is controlled by the integrated industrial panel PC. The modular build allows for the analyser to match your application and operational needs.
• The standard measuring range can be narrowed by a different calibration range or extended via internal dilution options. • Analogue and digital output options • Multiple stream analysis for up to 8 sample streams
There are many additional options available. Please contact Hach for more details.
EZ9501.99000001EZ Rack, Base Unit, for Reagents Only
Overview
EZ4000 Series Analysers are single-parameter titrators that achieve excellent precision and accuracy. Depending on the parameter of interest, the measuring range and the water matrix the EZ4000 Series may use either potentiometry i.e. acid-base, redox or precipitation titration, where the endpoint is determined by a change in a specific variable, or photocolorimetry, where colour change is used to determine the endpoint of the titration. The high precision dispenser for addition, robust peristaltic pumps for sampling and drain, and carefully designed liquid pathways all add up to the highest performance for industrial and environmental analysis needs.
Results you can rely on Smart automatic features for validation, priming and cleaning are embedded in the controller software and contribute to analytical performance, maximised uptime and negligible operator invervention. Sample lines and analysis vessel are cleaned with demineralised water to eliminate cross contamination between samples. Electronic and wet-chemical part of the analyser are strictly separated. A transparent door allows for instant visual inspection of the wet part.
Flexibility that meets your needs EZ Series Chloride Analysers come in an attractive, ergonomic mainframe with a compact footprint. All hardware is controlled by the integrated industrial panel PC. The modular build allows for the analyser to match your application and operational needs.
• The standard measuring range can be narrowed by a different calibration range or extended via internal dilution options. • Analogue and digital output options • Multiple stream analysis for up to 8 sample streams
There are many additional options available. Please contact Hach for more details.
Applications
Municipal Water
Ultrapure Water
Specifications
What's included?EZ4006 Chloride Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x mV electrode, 2 x empty 10L Reagent Containers with Fittings (for Cleaning & Buffer Solution) and 1 x empty 10L Reagent Dark Container with Fittings (for AgNO₃ Solution)
Number of sample streams1, 2, 4 or 8
Demineralised waterFor rinsing / dilution
Measurement methodPotentiometric titration with silver nitrate (AgNO₃), conform with standard method APHA 4500-Cl (D)
Documents & Downloads
Brochures
Brochure
Protect your investment and peace of mind - EZ Series Analyzer Service
EZ4000 Series Analysers are single-parameter titrators that achieve excellent precision and accuracy. Depending on the parameter of interest, the measuring range and the water matrix the EZ4000 Series may use either potentiometry i.e. acid-base, redox or precipitation titration, where the endpoint is determined by a change in a specific variable, or photocolorimetry, where colour change is used to determine the endpoint of the titration. The high precision dispenser for addition, robust peristaltic pumps for sampling and drain, and carefully designed liquid pathways all add up to the highest performance for industrial and environmental analysis needs.
Results you can rely on Smart automatic features for validation, priming and cleaning are embedded in the controller software and contribute to analytical performance, maximised uptime and negligible operator invervention. Sample lines and analysis vessel are cleaned with demineralised water to eliminate cross contamination between samples. Electronic and wet-chemical part of the analyser are strictly separated. A transparent door allows for instant visual inspection of the wet part.
Flexibility that meets your needs EZ Series Chloride Analysers come in an attractive, ergonomic mainframe with a compact footprint. All hardware is controlled by the integrated industrial panel PC. The modular build allows for the analyser to match your application and operational needs.
• The standard measuring range can be narrowed by a different calibration range or extended via internal dilution options. • Analogue and digital output options • Multiple stream analysis for up to 8 sample streams
There are many additional options available. Please contact Hach for more details.
Applications
Municipal Water
Ultrapure Water
Specifications
What's included?EZ4006 Chloride Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x mV electrode, 2 x empty 10L Reagent Containers with Fittings (for Cleaning & Buffer Solution) and 1 x empty 10L Reagent Dark Container with Fittings (for AgNO₃ Solution)
Number of sample streams1, 2, 4 or 8
Demineralised waterFor rinsing / dilution
Measurement methodPotentiometric titration with silver nitrate (AgNO₃), conform with standard method APHA 4500-Cl (D)
Documents & Downloads
Brochures
Brochure
Protect your investment and peace of mind - EZ Series Analyzer Service
PrecisionBetter than 2% full scale range for standard test solutions
ValidationAutomatic; frequency freely programmable
Flow rate100 - 300 mL/min
Warranty24 months
Sample temperature10 - 30 °C
ParameterChloride
Instrument airDry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
ModelEZ4006
CertificationsCE compliant / UL certified
InterferencesBromide, sulphide, iodide ions may interfere. Ferricyanide causes high results and must be removed. Chromate and dichromate interfere and should be reduced to chromic state or removed. Ferric iron interferes if present in an amount substantially higher than the amount of chloride. Chromic ion, ferrous ion and phosphate do not interfere. Fats, oil, proteins, surfactants and tar.
PrecisionBetter than 2% full scale range for standard test solutions
ValidationAutomatic; frequency freely programmable
Flow rate100 - 300 mL/min
Warranty24 months
Sample temperature10 - 30 °C
ParameterChloride
Instrument airDry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
ModelEZ4006
CertificationsCE compliant / UL certified
InterferencesBromide, sulphide, iodide ions may interfere. Ferricyanide causes high results and must be removed. Chromate and dichromate interfere and should be reduced to chromic state or removed. Ferric iron interferes if present in an amount substantially higher than the amount of chloride. Chromic ion, ferrous ion and phosphate do not interfere. Fats, oil, proteins, surfactants and tar.