Large, high-resolution, transreflective display provides optimal viewing resolution in any lighting condition. Guided calibration procedures in 19 languages minimise complexity and reduce operator error. Password-protected SD card reader offers a simple solution for data download and transfer. Visual warning system provides critical alerts.
Wide variety of communication options
Utilise two to five analog outputs to transmit primary and secondary values for each sensor, or integrate Hach sensors and analysers into Hart networks.
Applications
Municipal Water
Specifications
Power requirements (Hz)50/60 Hz
Display size48 x 68 mm
Power Requirements (Voltage)100 - 240 V AC, 24 V DC
Warranty24 months
Operating Temperature Range
Documents & Downloads
Brochures
Brochure
Care Advantage Flyer: Complete service solution for all your instruments
The SC200 controller allows the use of digital and analogue sensors, either alone or in combination, to provide compatibility with Hach's broad range of sensors, eliminating the need for dedicated, parameter-specific controllers.
Ease of use and confidence in results
Large, high-resolution, transreflective display provides optimal viewing resolution in any lighting condition. Guided calibration procedures in 19 languages minimise complexity and reduce operator error. Password-protected SD card reader offers a simple solution for data download and transfer. Visual warning system provides critical alerts.
Wide variety of communication options
Utilise two to five analog outputs to transmit primary and secondary values for each sensor, or integrate Hach sensors and analysers into Hart networks.
Applications
Municipal Water
Specifications
Power requirements (Hz)50/60 Hz
Display size48 x 68 mm
Power Requirements (Voltage)100 - 240 V AC, 24 V DC
Warranty24 months
Operating Temperature Range
Documents & Downloads
Brochures
Brochure
Care Advantage Flyer: Complete service solution for all your instruments
RelaysFour electromechanical SPDT (Form C) contacts, 1200 W, 5 A
Display resolution240 x 160 pixels
Weight1.7 kg
Storage conditions-20 - 70 °C
Dimensions (H x W x D)144 mm x 144 mm x 181 mm
Conduit openings1/2“ NPT conduit
Relay: Operational modePrimary or secondary measurement, calculated value (dual channel only) or timer
Enclosure waterproof ratingIP66 / NEMA 4X
Analogue outputsTwo (five with optional expansion module) to isolated current outputs, max 550 Ω , accuracy: ± 0.1% of FS (20mA) at 25 °C, ± 0.5% of FS over -20 °C to 60 °C range Operational mode: measurement or calculated value
DisplayGraphic dot matrix LCD with LED backlighting, transreflective
Electrical CertificationsEMC CE compliant for conducted and radiated emissions: - CISPR 11 (Class A limits) - EMC Immunity EN 61326-1 (Industrial limits) Safety cETLus safety mark for: - General Locations per ANSI/UL 61010-1 & CAN/CSA C22.2. No. 61010-1 - Hazardous Location Class I, Division 2, Groups A,B,C & D (Zone 2, Group IIC) per FM 3600 / FM 3611 & CSA C22.2 No. 213 M1987 with approved options and appropriately rated Class I, Division 2 or Zone 2 sensors cULus safety mark - General Locations per UL 61010-1 & CAN/CSA C22.2. No. 61010-1
RelaysFour electromechanical SPDT (Form C) contacts, 1200 W, 5 A
Display resolution240 x 160 pixels
Weight1.7 kg
Storage conditions-20 - 70 °C
Dimensions (H x W x D)144 mm x 144 mm x 181 mm
Conduit openings1/2“ NPT conduit
Relay: Operational modePrimary or secondary measurement, calculated value (dual channel only) or timer
Enclosure waterproof ratingIP66 / NEMA 4X
Analogue outputsTwo (five with optional expansion module) to isolated current outputs, max 550 Ω , accuracy: ± 0.1% of FS (20mA) at 25 °C, ± 0.5% of FS over -20 °C to 60 °C range Operational mode: measurement or calculated value
DisplayGraphic dot matrix LCD with LED backlighting, transreflective
Electrical CertificationsEMC CE compliant for conducted and radiated emissions: - CISPR 11 (Class A limits) - EMC Immunity EN 61326-1 (Industrial limits) Safety cETLus safety mark for: - General Locations per ANSI/UL 61010-1 & CAN/CSA C22.2. No. 61010-1 - Hazardous Location Class I, Division 2, Groups A,B,C & D (Zone 2, Group IIC) per FM 3600 / FM 3611 & CSA C22.2 No. 213 M1987 with approved options and appropriately rated Class I, Division 2 or Zone 2 sensors cULus safety mark - General Locations per UL 61010-1 & CAN/CSA C22.2. No. 61010-1